|Table of Contents|

Numerical Studies on Temperature Distribution Analysis and Treatment Efficacy Evaluation for High Intensity Focused Ultrasound Therapy(PDF)

《南京师大学报(自然科学版)》[ISSN:1001-4616/CN:32-1239/N]

Issue:
2017年01期
Page:
144-
Research Field:
·物理学·
Publishing date:

Info

Title:
Numerical Studies on Temperature Distribution Analysis and Treatment Efficacy Evaluation for High Intensity Focused Ultrasound Therapy
Author(s):
Su Huidan1Dai Sijie2Guo Gepu1Ma Qingyu1
(1.School of Physics and Technology,Nanjing Normal University,Nanjing 210023,China)(2.Honors College,Nanjing Normal University,Nanjing 210023,China)
Keywords:
HIFU therapystructure coefficienttemperature distributionefficacy evaluation
PACS:
O426.2
DOI:
10.3969/j.issn.1001-4616.2017.01.021
Abstract:
High intensity focused ultrasound(HIFU)is a non-invasive therapeutic medical treatment modality,which can destroy tumor tissues without damaging to the surrounding normal tissues and has less possibility of cancer metastasis. The key issue of accurate HIFU therapeutic dosage control is how to monitor the efficacy in the process of HIFU treatment. Based on HIFU localized tissue heating,the model of HIFU treatment is established using the finite element method(FEM),and the numerical simulations of the acoustic and temperature fields at HIFU focal region are developed. The influences of transducer parameters,acoustic power and treatment time on temperature distribution at the focal region are discovered. It is proved that,with the increasing treatment time,the ellipsoidal focal region forms gradually with a rising temperature distribution. Once reaching HIFU treatment efficacy,the temperature of the focal region(in radial±0.4 mm)reaches 343 K(70 ℃),with the decrease of the structure coefficient(radius/focal length)of the transducer,the required treatment time increases,producing a narrower ellipsoidal focal region with an enhanced ratio between the radial width and axial width. Therefore,the simulations of temperature field for the transducer with a known structure coefficient in this study provide a new method for temperature estimation,efficacy evaluation and dose control in the HIFU therapy.

References:

[1] HUTCHINSON L. Treatment modalities:HIFU is effective for unresectable HCC[J]. Nature reviews clinical oncology,2011,8(7):385-385.
[2]KENNEDY J E. High-intensity focused ultrasound in the treatment of solid tumors[J]. Nature reviews cancer,2005,5(4):321-327.
[3]何向东,冯少彤. 高强度聚焦超声温度场的有限差分模拟[J]. 南京师大学报(自然科学版),2015,38(4):106-112.
[4]陈轩泽,马青玉,杨玲,等. 离体猪肝组织电阻抗的温度特性研究[J]. 南京师大学报(自然科学版),2012,35(2):32-38.
[5]XING Y,LU X,PUA E C,et al. The effect of high intensity focused ultrasound treatment on metastases in a murine melanoma model[J]. Biochemical and biophysical research communications,2008,375(4):645-650.
[6]CLARKE R L,TER HAAR G R. Temperature rise recorded during lesion formation by high-intensity focused ultrasound.[J]. Ultrasound in medicine and biology,1997,23(2):299-306.
[7]姜立新,胡兵. 高强度聚焦超声作用下的组织温度场测量技术的研究进展[J]. 声学技术,2006,25(1):43-47.
[8]JOLESZ F A,MCDANNOLD N. Current status and future potential of MRI-guided focused ultrasound surgery[J]. Journal of magnetic resonance imaging,2008,27(2):391-399.
[9]YE G,SMITH P P,NOBLE J A. Model-based ultrasound temperature visualization during and following Hifu exposure[J]. Ultrasound in medicine and biology,2010,36(2):234-249.
[10]QIAN Z W,YU J,FEI X,et al. Noninvasive temperature monitoring in Hifu clinical uses[J]. International symposium on therapeutic ultrasound proceedings,2007,911(1):251-257.
[11]DANIELS M J,VARGHESE T,MADSEN E L,et al. Non-invasive ultrasound-based temperature imaging for monitoring radiofrequency heating-Phantom results[J]. Physics in medicine and biology,2007,52(16):4 827-4 843.
[12]MA Y,ZHANG D,GONG XF,et al. Noninvasive temperature estimation by detecting echo-strain change including thermal expansion[J]. Chinese physics,2007,16(9):2 745-2 751.
[13]ANAND A,KACZKOWSKI P J. A model-based noninvasive temperature estimation technique for HIFU therapy monitoring using backscattered ultrasound[J]. Journal of the acoustical society of America,2004,115(5):2 490-2 490.
[14]BLACKSTOCK D T. Fundamentals of physical acoustics[M]. New York:John Wiley & Sons Inc.,2000:476-488.
[15]范庭波. 高强度聚焦超声非线性声场和组织损伤形成的扫描模式比较研究[D]. 南京:南京大学,2012.
[16]PENNES H H. Analysis of tissue and arterial blood temperatures in the resting human forearm[J]. Journal of applied physiology,1998,85(1):5-34.
[17]胡兵,姜立新,黄瑛. 用于高强度聚焦超声热消融的PAA模块:模块的制作及声学参数的测量(英文)[J]. 声学技术,2006,25(6):613-616.
[18]KREIDER W,YULDASHEV P V,SAPOZHNIKOV O A,et al. Characterization of a multi-element clinical HIFU system using acoustic holography and nonlinear modeling[J]. IEEE transactions on ultrasonics ferroelectrics and frequency control,2013,60(8):1 683-1 698.
[19]ZHANG C B,CAO H L,LI Q,et al. Enhancement effect of ultrasound-induced microbubble cavitation on branched polyethylenimine-mediated VEGF(165)transfection with varied N/P ratio.[J]. Ultrasound in medicine and biology,2013,39(1):161-171.
[20]MYERS M R,Soneson J E. Temperature modes for nonlinear Gaussian beams[J]. Journal of the acoustical society of America,2009,126(1):425-433.
[21]SONESON J E,MYERS M R. Gaussian representation of high-intensity focused ultrasound beams[J]. Journal of the acoustical society of America,2007,122(5):2 526-2 531.
[22]SONESON J E,MYERS M R. Thresholds for nonlinear effects in high-intensity focused ultrasound propagation and tissue heating[J]. IEEE transactions on ultrasonics ferroelectrics and frequency control,2010,57(11):2 450-2 459.
[23]BESSONOVA O,WILKENS V. Non-invasive estimation of temperature using diagnostic ultrasound during high intensity focused ultrasound therapy[J]. Journal of the acoustical society of America,2013,134(5):4 213-4 213.
[24]CHEN T,FAN T,ZHANG W,et al. Acoustic characterization of high intensity focused ultrasound fields generated from a transmitter with a large aperture[J]. Journal of applied physics,2014,115(11):114 902-1-114 902-7.

Memo

Memo:
-
Last Update: 1900-01-01